0
Scientific Article   |    
In Vivo Deterioration of Tibial Baseplate Locking Mechanisms in Contemporary Modular Total Knee Components
Gerard A. Engh, MD; Smain Lounici, MS; Anand R. Rao, BS; Matthew B. Collier, MS
The Journal of Bone & Joint Surgery.  2001; 83:1660-1665 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: The results of recent studies documenting the backside wear of polyethylene inserts retrieved from total knee implants call into question the stability of the locking mechanisms of modular tibial components. Wear of the metal tibial baseplate suggests that the capture mechanisms of some modular fixed-bearing tibial components do not adequately restrict in vivo motion of the insert. The purposes of this study were (1) to present a method for evaluating locking-mechanism stability and (2) to investigate the stability of modular tibial components after an interval in vivo.

Methods: We measured the anteroposterior and mediolateral motion between the polyethylene insert and the tibial tray in a variety of modular total knee tibial components. A uniaxial mechanical testing machine was used to evaluate the stability of ten unimplanted components (control group), fifteen implants obtained from patients who were undergoing revision total knee arthroplasty (revision group), and fifteen devices retrieved post mortem (autopsy group). We applied loads along the anteroposterior and mediolateral axes of the tibial component and recorded the maximum insert displacement that occurred. From this value, we calculated an insert-motion index, the magnitude of a two-dimensional vector that represented the total motion in the transverse plane.

Results: For the control group, the mean insert-motion index was 64 ± 13 m (range, 6 to 157 m); for the revision group, it was 341 ± 51 m (range, 104 to 718 m); and for the autopsy group, it was 380 ± 45 m (range, 122 to 657 m). The insert-motion index for the control group was significantly lower than that for the revision group (p = 0.001) or autopsy group (p < 0.001).

Conclusions: Motion between the polyethylene insert and the metal baseplate in contemporary modular tibial designs increases after a period of in vivo loading.

Clinical Relevance: Although there are several advantages to the use of modular tibial components, these ad-vantages must be weighed against the disadvantage of backside wear debris secondary to motion of the modular insert. Debris from backside wear combined with wear from the articular side might account for the increasing prevalence of osteolysis since modular components have become widely used.

Figures in this Article
    Polyethylene wear debris from the bearing surfaces of contemporary modular total knee components is widely recognized as the primary cause of osteolysis after total knee arthroplasty. To our knowledge, osteolysis was not reported as a clinical problem with the first generation of one-piece tibial components. Only after the introduction of modular polyethylene inserts in the mid-1980s was failure of knee replacements due to osteolysis recognized as a major clinical problem1,2. In recent years, researchers also have observed wear on the backside of retrieved polyethylene inserts3-5 (Figs. 1-A and 1-B). Perhaps the combination of articulating surface wear and backside wear has produced a greater volume of debris, which has caused the increased occurrence of osteolysis observed with the use of modular implants.
    The presence of backside polyethylene wear on retrieved components indicates that the locking mechanism of fixed-bearing tibial trays may not adequately eliminate motion between the modular elements2,3. Although motion of the insert has been documented in a laboratory setting4, to our knowledge the in vivo increase or reduction of the insert motion of contemporary modular components has yet to be reported. In the present study, we hypothesized that the motion that is inherent in implants that have a locking mechanism would increase over time with in vivo service and could contribute to the generation of wear debris from the metal baseplate and the backside of the polyethylene insert. Our purposes were (1) to present a method for evaluating the stability of the locking mechanism of new and retrieved modular tibial components, and (2) to test the stability of modular components that were retrieved post mortem and at revision of total knee replacements to determine whether the stability of the locking mecha-nism deteriorated after in vivo loading.
     
    Anchor for JumpAnchor for Jump
    +Fig. 1-A:This snap-fit tibial baseplate shows an extensive amount of stippling in a radial pattern.
     
    Anchor for JumpAnchor for Jump
    +Fig. 1-B:On the undersurface of the corresponding tibial polyethylene insert, the manufacturer’s product-description letters and numbers are wearing away.
     
    Anchor for JumpAnchor for Jump
    +Fig. 2:The polyethylene insert and the -baseplate were mounted in separate frames on a materials testing system. -Displacement was measured with extensometers.
     
    Anchor for JumpAnchor for Jump
    +Fig. 3:The insert motion was the -displacement measured between the insert and tray before resistance from the baseplate -locking mechanism was encountered (in other words, while the materials testing system detected a constant load).
     
    Anchor for JumpAnchor for JumpTABLE I:  Modular Tibial Designs* Included in the Study
    *The Anatomic Modular Knee is manufactured by DePuy, Warsaw, Indiana; Axiom, by Orthomet, Minneapolis, Minnesota; Duracon, by Howmedica, -Rutherford, New Jersey; Genesis, by Richards, Memphis, Tennessee; Insall-Burstein II, by Zimmer, Warsaw, Indiana; Kinematic II, by How-medica, Rutherford, New Jersey; Maxim, by Biomet, Warsaw, Indiana; Miller-Galante II, by Zimmer, Warsaw, Indiana; Ortholoc, by Wright Medical Technology, Arlington, Tennessee; Press Fit Condylar, by Johnson and Johnson, Raynham, Massachusetts; Synatomic, by DePuy, Warsaw, Indiana.
    Control Group (10 Specimens, 10 Designs)Revision Group (15 Specimens, 10 Designs)Autopsy Group (15 Specimens, 4 Designs)
    Anatomic Modular Knee1 Anatomic Modular Knee7 Anatomic Modular Knee
    Axiom1 Axiom
    Duracon2 Duracon
    Genesis2 Genesis
    Insall-Burstein II2 Insall-Burstein II1 Insall-Burstein II
    Kinematic II1 Kinematic II
    Maxim1 Maxim
    Miller-Galante II1 Miller-Galante II
    Ortholoc2 Ortholoc
    Press Fit Condylar2 Press Fit Condylar4 Press Fit Condylar
    3 Synatomic
    We tested forty tibial components from a variety of modular, fixed--bearing total knee designs. The components were categorized into three groups: unimplanted components, implants obtained during revision procedures, and devices -retrieved post mortem (Table I). Thirty-two components were cruciate-retaining, and eight were cruciate-substituting. Of the unimplanted components, only the Insall-Burstein-II -im-plant was cruciate-substituting. Seven of the retrieved components were cruciate-substituting, including three Insall-Burstein II implants (Zimmer, Warsaw, Indiana), two Genesis implants (Richards, Memphis, Tennessee), one Or-tholoc implant (Wright Medical Technology, Arlington, Tennessee), and one Press Fit Condylar implant (Johnson and Johnson, Raynham, Massachusetts).
    The group of unimplanted components served as our controls. The -control group consisted of ten midsized components, each with a different -locking-mechanism design (Table I). For each component, we selected a polyethylene-insert thick-ness that most closely approximated a tibial component thickness of 10 mm when combined with the tibial baseplate.
    The revision group consisted of fifteen modular tibial components that had been retrieved from fifteen patients during revision total knee arthroplasty. We used our database to identify the modular tibial components that had the same locking mechanism as the controls and that had the longest in vivo duration for each design and then selected those revision implants from our retrieval laboratory. One or two implants of each locking-mechanism design were selected. All fractured inserts and those with apparent damage to the locking mechanism were excluded from the study.
    Eleven of the revision components were from women, and four were from men. At the time of implant removal, the average patient age was 64.3 years (range, thirty-six to eighty years) and the average patient weight was 90.3 kg (range, 52.2 to 145.2 kg). The mean time that the implants were in situ was forty-four months (range, two to eighty-eight months). The preoperative diagnosis that necessitated the primary total knee arthroplasty was osteoarthritis for thirteen patients and posttraumatic arthritis for two. The revision total knee arthroplasty was performed for a variety of reasons, including osteolysis (four patients); a deep infection (three); a loose cemented tibial component (two); and knee instability, loose cementless tibial and femoral components, a fracture of a femoral component, oversized tibial and femoral components, arthrofibrosis, and patellar subluxation (one patient each).
    The autopsy group consisted of fifteen modular tibial components harvested post mortem along with the supporting soft tissue and bone. The specimens included components of four different tibial designs, including three of the designs that were tested in the control group. Seven implants were from five women, and eight were from five men. The average age of the patients at the time of death was seventy-eight years (range, sixty-nine to eighty-four years), and the average weight was 80.4 kg (range, 55.8 to 131.6 kg). The average time that the implants were in situ was eighty-seven months (range, one to 158 months). All of the total knee arthroplasties in the autopsy group had been performed because of osteoarthritis, with the exception of two that had been performed bilaterally in one patient because of rheumatoid arthritis. While the patients were living, the average Hospital for Special Surgery knee-rating score6 was 91.5 points (range, 85 to 98 points) for four knees, and the average Knee Society Score7 was 89.7 points (range, 73 to 99 points) for nine knees. All of the patients had reported little or no pain at the final follow-up evaluation.
    All but one insert retrieved at revision had been separated from the tibial baseplate during the surgical procedure. We tested this specimen and three autopsy implants both before and after disassembling the components to determine whether disassembly affected motion of the insert relative to the baseplate. For these four components, only the test results after the reassembly were considered when we compiled or compared the experimental data.
    The anteroposterior and mediolateral motion between the polyethylene insert and the baseplate of the tibial tray of each control and retrieved implant was measured with use of the following protocol. The control implants were assembled according to the manufacturer’s instructions. The specimens in each group were preconditioned for two weeks in a bath of saline solution at 37°C to simulate in vivo physiological conditions, which caused the polyethylene to swell and the locking mechanism to tighten8. Each tibial baseplate, or proximal part of the tibia for the autopsy specimens, was secured inside a rectangular solid mold with the use of set-screws so that the anteroposterior, mediolateral, and superior-inferior axes of the tray were aligned with the dimensions of the mold. The mold was then filled with a quick-setting acrylic (Bondo body filler; Bondo/Mar-Hyde, Atlanta, Georgia), which, after hardening, yielded a fixture geometry that allowed for the consistent attachment of the various tibial baseplate designs to the testing machine. The baseplate and the polyethylene insert were mounted separately on two metal frames (Fig. 2), which were then mounted vertically on a materials testing machine (MTS Systems, Eden Prairie, Minnesota) with a dovetail system. Metal brackets were affixed to the polyethylene insert and the acrylic mold that housed the baseplate with the use of hot glue. The brackets provided reference points for an extensometer to record the anteroposterior and the mediolateral displacement between the insert and the baseplate.
    At a fixed rate of 10 N/sec, we applied a compressive load of 100 N to the frame that held the polyethylene insert, which induced downward motion of the polyethylene insert relative to the stationary tibial baseplate. After attaining 100 N of compression, the system was loaded to 100 N of tension at the same loading rate, displacing the insert upward relative to the baseplate. This loading profile was applied for at least one additional cycle, and the resulting plot was inspected to ensure reproducibility of the output. The fixture was then rotated 90° about the superior-inferior axis of the component to align the alternate axis with the direction of loading. The above test was repeated to determine the permissible displacement in the plane of the tibial baseplate. The recorded parameters during each test of anteroposterior or mediolateral insert displacement were time (in seconds), load (in newtons), and exten-someter displace-ment (in micrometers).
    This protocol ensured that the total slack of the locking mechanism was recorded, regardless of the initial position of the insert relative to the baseplate. We defined insert motion as the displacement that occurred between the insert and the baseplate with no mechanical resistance (Fig. 3). In order to characterize the stability of the locking mechanism, we then calculated the -insert-motion index, the magnitude of a two-dimensional vector that represented the total motion in the transverse plane, for each specimen with use of the equation: Insert-Motion Index = AP2+ML2, where AP is the insert motion in the anteroposterior direction and ML is the insert motion in the mediolateral direction.
    We used analysis of variance with a Tukey post hoc test to analyze the differences in the insert-motion index among the three groups. A detectable difference of 200 m between groups was considered clinically relevant. A sample power analysis revealed that, for the standard deviation levels for the -anteroposterior and mediolateral motions, a sample size of eight would lead to a power of 0.95 (alpha = 0.05). Statistical relationships between the insert-motion- index and other variables were assessed with the use of the Mann-Whitney U- test or the Spearman rho test.
    The mean insert-motion index (and standard error) was 64 ± 13 m (range, 6 to 157 m) for the ten tibial components in the control group, 341 ± 51 m (range, 104 to 718 m) for the fifteen components in the revision group, and 380 ± 45 m (range, 122 to 657 m) for the fifteen autopsy specimens. The insert-motion index of the control group was significantly lower than that of the revision group (p = 0.001, analysis of variance, Tukey post hoc test) and that of the autopsy group (p < 0.001, analysis of variance, Tukey post hoc test). There was no significant difference between the insert-motion indices of the revision and autopsy groups (p = 0.8, analysis of variance, Tukey post hoc test).
    Included among the two retrieval groups were twenty-seven specimens of the ten designs tested in the control group. The insert-motion index for each of these retrieved specimens exceeded that of its control counterpart by an -average of 273 m (range, 43 to 696 m). The insert-motion indices were similar in the left and right components from three patients in the autopsy group who had received the same type of total knee implant bilaterally and in whom the in situ durations of the two implants had been equivalent. The values for the two implants in each of these three patients were 122 m compared with 231 m, 465 m compared with 488 m, and 561 m compared with 625 m. In contrast, the test results varied among components of the same design that had been implanted in different patients.
    In the revision and autopsy groups, no correlations were found between the insert-motion index and the in situ duration (p = 0.8, Spearman rho test), the patient’s weight (p = 0.9, Spearman rho test), or the patient’s gender (p = 0.6, Mann-Whitney U test). In situ duration was significantly longer for the autopsy group than for the revision group (p < 0.003, Mann-Whitney U test). In addition, no association was found between the mechanism of failure that had necessitated revision and the insert-motion index. The four specimens revised because of osteolysis had highly variable insert-motion indices (121, 178, 276, and 718 m) relative to the general population of retrieval specimens.
    The comparison of insert-motion indices before and after disassembly of the modular tibial component demonstrated variable changes in insert stability. After reassembly, one revision and one autopsy implant had increased insert motion (+19 and +39 m, respectively), and two autopsy implants had decreased insert motion (-24 and -69 m). All four inserts had greater motion than their respective control specimens, both before and after modular disassembly.
    The results of our study demonstrate that the instability inherent in contemporary modular tibial tray locking mechanisms increases with in vivo physiological loading. The elimination of all motion between modular parts that have markedly different moduli of elasticity is not possible. We believe that after implantation a competition phenomenon occurs between the stability of the modular tibial tray locking mechanism and the stresses acting upon it. The ability of the locking mechanism to counteract repetitively applied stresses is affected by the initial insert motion, the in situ duration, and the stresses transmitted at the modular interface. We speculate that the instability of the locking mechanism begins with deformation of the polyethylene and then gradually increases as wear of the polyethylene insert further alters the fit within its metal housing.
    Parks et al.4 previously reported the interface motion of nine types of unimplanted modular tibial trays. Using the same implant designs, we confirmed their finding that motion can occur in assembled tibial components with each capture-mechanism design. The magnitudes of motion under 100-N and 400-N loads reported by Parks et al. represented a combination of polyethylene deformation and insert motion. We chose to isolate insert motion from polyethylene deformation by identifying the displacement that occurred before resistance from the locking mechanism was encountered (Fig. 3). Therefore, the lower values in our study are inherent to the nature of our loading protocol. The insert-motion index expresses the magnitude of the motion vectors in the transverse plane, a Cartesian combination of both anteroposterior and mediolateral motion. We believe that this is a valid method for testing the locking mechanisms of new and retrieved modular fixed-bearing inserts and is more representative of motion that might occur under minimal in vivo loading conditions.
    The amount of motion that occurs with in vivo loading is not necessarily represented by the amount measured in the autopsy and retrieval specimens. We did not account for axial or torsional loads or for other factors that interact with the stability of the locking mechanism in vivo. Axial loads due to muscle forces and weight-bearing would be expected to affect insert motion in vivo, but we did not apply such loads in the mechanical testing. For ease of measurement, we recorded insert motion that was parallel to the anteroposterior and mediolateral axes of the baseplate under loads directed parallel to these axes, although characteristic wear patterns on explanted tibial tray baseplates confirm that rotational motion also occurs at the interface in many designs (Fig. 1-A). We acknowledge that rotation in the same plane also occurs, as does motion perpendicular to the tibial component.
    We speculated that removal of the insert might alter the stability of the modular tibial components. However, we found that the insert-motion indices before and after disassembly were comparable, suggesting that insert removal did not appreciably alter insert stability.
    Although only four different designs were available in the postmortem group, the test results for these implants were similar to the results for the wide variety of revision implants tested.
    No difference in insert motion was found between the revision and autopsy specimens. Perhaps the shorter length of time for revision components to exhibit insert motion equivalent to that of the autopsy specimens was associated with the conditions that had necessitated revision surgery and the younger average age of the patients in the revision group.
    We had speculated that the four implants revised because of osteolysis would have more insert motion than the other revision components. However, we did not find a relationship between the reason for component revision and insert stability. We also had thought that the demands on the stability of the locking mechanism might be greater for the more conforming fixed-bearing implants, such as the posterior-stabilized and varus-valgus constrained implants. In theory, a higher degree of articular constraint would transmit higher loads to the insert-baseplate interface in these designs. However, we did not find more insert motion in the six posteriorly stabilized components.
    Tibial tray-locking mechanisms come in a variety of designs that could lead to some differences in insert stability4. We found that tongue-in-groove components blocked motion least effectively in the direction of the tongue and groove. Therefore, Anatomic Modular Knee implants demonstrated greater anteroposterior motion, and Insall-Burstein-II implants showed greater mediolateral motion. In the control group, the snap-fit implants, such as the Press Fit Condylar and Miller-Galante-II components, were marginally more stable than the tongue-in-groove components. However, the retrieved snap-fit components demonstrated overall motion equivalent to that of the tongue-in-groove components, but with greater variability in the direction of motion. Components with a partial capture wall, such as the Genesis and the Duracon implants, tended to allow motion in the direction without a barrier to insert motion.
    Caution is needed when the femoral articulating surface and the insert-tray interface are being compared. A few hundred micrometers of motion at the tray-insert interface is much smaller than the motion that occurs at the tibiofemoral articulation as the knee flexes through an arc of >90°. However, the articular side has a highly polished surface, whereas the baseplates of most modular tibial components do not. Certainly, two relatively rough surfaces moving against each other with loads exceeding three times body weight will generate substantial amounts of debris. Our findings warrant additional studies to examine the instability of locking mechanisms and the correlations with the extent of wear of the counterfaces.
    Modular tibial components have a number of advantages. Inventory is reduced when variable thickness polyethylene inserts can be combined with tibial baseplates of different sizes. Positioning and insertion of the components is far easier with use of modular implants. Cementless fixation of the tibial tray is enhanced because screw augmentation is allowed. Revision surgery is easier because of improved access to the tibial fixation interface with removal of the modular insert. Insert exchange provides an alternative to complete revision if polyethylene wear becomes a clinical problem. However, until problems with modularity are resolved, these relative advantages of modular components need to be considered against the consequences of more wear debris.
    The results of this study indicate that locking-mechanism designs for modular implants need to be improved. An acceptable amount of motion of modular components has not been established. The mechanical stability of contemporary modular components should be documented by benchmark testing under repetitive physiological loads and by additional testing of retrieved modular implants. If insert motion cannot be reduced to an acceptable level with the use of modular components, the orthopaedic community should consider -establishing International Organization for Standardization standards for the surface finish of modular elements. Alternatives to reduce wear debris, such as nonmodular implants and polished tibial baseplates, should be considered until the stability of modular components can be ensured.
    Peters PC, Engh GA, Dwyer KA,Vinh TN. Osteolysis after total knee arthroplasty without cement. J Bone Joint Surg Am,1992;74: 864-76. 74864  1992  [PubMed]
     
    Engh GA, Dwyer KA,Hanes CK. Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses. J Bone Joint Surg Br,1992;74: 9-17. 749  1992  [PubMed]
     
    Wasielewski RC, Parks N, Williams I, Surprenant H, Collier JP,Engh G. Tibial insert undersurface as a contributing source of polyethylene wear debris. Clin Orthop,1997;345: 53-9. 34553  1997  [PubMed]
     
    Parks NL, Engh GA, Topoleski LD,Emperado J. The Coventry Award. Modular tibial insert micromotion. A concern with contemporary knee implants. Clin Orthop,1998;356: 10-5. 35610  1998  [PubMed][CrossRef]
     
    Gabriel SM, Dennis DA, Honey MJ,Scott RD. Polyethylene wear on the distal tibial insert surface in total knee arthroplasty. Knee,1998;5: 221-8. 5221  1998  [CrossRef]
     
    Insall JN, Ranawat CS, Aglietti P,Shine J. A comparison of four models of total knee--replacement prostheses. J Bone Joint Surg Am,1976;58: 754-65. 58754  1976  [PubMed]
     
    Insall JN, Dorr LD, Scott RD,Scott WN. Rationale of the Knee Society clinical rating system. Clin Orthop,1989;248: 13-4. 24813  1989  [PubMed]
     
    Clarke IC, Starkebaum W, Hosseinian A, McGuire P, Okuda R, Salovey R,Young R. Fluid--sorption phenomena in sterilized polyethylene acetabular prostheses. Biomaterials,1985;6: 184-8. 6184  1985  [PubMed][CrossRef]
     

    Submit a comment

    Anchor for JumpAnchor for Jump
    +Fig. 1-A:This snap-fit tibial baseplate shows an extensive amount of stippling in a radial pattern.
    Anchor for JumpAnchor for Jump
    +Fig. 1-B:On the undersurface of the corresponding tibial polyethylene insert, the manufacturer’s product-description letters and numbers are wearing away.
    Anchor for JumpAnchor for Jump
    +Fig. 2:The polyethylene insert and the -baseplate were mounted in separate frames on a materials testing system. -Displacement was measured with extensometers.
    Anchor for JumpAnchor for Jump
    +Fig. 3:The insert motion was the -displacement measured between the insert and tray before resistance from the baseplate -locking mechanism was encountered (in other words, while the materials testing system detected a constant load).
    Anchor for JumpAnchor for JumpTABLE I:  Modular Tibial Designs* Included in the Study
    *The Anatomic Modular Knee is manufactured by DePuy, Warsaw, Indiana; Axiom, by Orthomet, Minneapolis, Minnesota; Duracon, by Howmedica, -Rutherford, New Jersey; Genesis, by Richards, Memphis, Tennessee; Insall-Burstein II, by Zimmer, Warsaw, Indiana; Kinematic II, by How-medica, Rutherford, New Jersey; Maxim, by Biomet, Warsaw, Indiana; Miller-Galante II, by Zimmer, Warsaw, Indiana; Ortholoc, by Wright Medical Technology, Arlington, Tennessee; Press Fit Condylar, by Johnson and Johnson, Raynham, Massachusetts; Synatomic, by DePuy, Warsaw, Indiana.
    Control Group (10 Specimens, 10 Designs)Revision Group (15 Specimens, 10 Designs)Autopsy Group (15 Specimens, 4 Designs)
    Anatomic Modular Knee1 Anatomic Modular Knee7 Anatomic Modular Knee
    Axiom1 Axiom
    Duracon2 Duracon
    Genesis2 Genesis
    Insall-Burstein II2 Insall-Burstein II1 Insall-Burstein II
    Kinematic II1 Kinematic II
    Maxim1 Maxim
    Miller-Galante II1 Miller-Galante II
    Ortholoc2 Ortholoc
    Press Fit Condylar2 Press Fit Condylar4 Press Fit Condylar
    3 Synatomic
    Peters PC, Engh GA, Dwyer KA,Vinh TN. Osteolysis after total knee arthroplasty without cement. J Bone Joint Surg Am,1992;74: 864-76. 74864  1992  [PubMed]
     
    Engh GA, Dwyer KA,Hanes CK. Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses. J Bone Joint Surg Br,1992;74: 9-17. 749  1992  [PubMed]
     
    Wasielewski RC, Parks N, Williams I, Surprenant H, Collier JP,Engh G. Tibial insert undersurface as a contributing source of polyethylene wear debris. Clin Orthop,1997;345: 53-9. 34553  1997  [PubMed]
     
    Parks NL, Engh GA, Topoleski LD,Emperado J. The Coventry Award. Modular tibial insert micromotion. A concern with contemporary knee implants. Clin Orthop,1998;356: 10-5. 35610  1998  [PubMed][CrossRef]
     
    Gabriel SM, Dennis DA, Honey MJ,Scott RD. Polyethylene wear on the distal tibial insert surface in total knee arthroplasty. Knee,1998;5: 221-8. 5221  1998  [CrossRef]
     
    Insall JN, Ranawat CS, Aglietti P,Shine J. A comparison of four models of total knee--replacement prostheses. J Bone Joint Surg Am,1976;58: 754-65. 58754  1976  [PubMed]
     
    Insall JN, Dorr LD, Scott RD,Scott WN. Rationale of the Knee Society clinical rating system. Clin Orthop,1989;248: 13-4. 24813  1989  [PubMed]
     
    Clarke IC, Starkebaum W, Hosseinian A, McGuire P, Okuda R, Salovey R,Young R. Fluid--sorption phenomena in sterilized polyethylene acetabular prostheses. Biomaterials,1985;6: 184-8. 6184  1985  [PubMed][CrossRef]
     
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Related Audio and Videos
    PubMed Articles
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    03/07/2012
    KY - University of Louisville Dept. of Orthopaedic Surgery
    01/04/2012
    PA - Penn State Milton S. Hershey Medical Center - Dept. of Orthopaedics & Rehabilitation
    05/18/2012
    NH - Concord Orthopaedics
    03/22/2012
    IL - Midwest Orthopaedics at Rush