Tibial Interface Wear in Retrieved Total Knee Components and Correlations with Modular Insert Motion
Anand R. Rao, BS; Gerard A. Engh, MD; Matthew B. Collier, MS; Smain Lounici, MS
View Disclosures and Other Information
Investigation performed at the Anderson Orthopaedic Research Institute, Alexandria, Virginia

Anand R. Rao, BS
Gerard A. Engh, MD
Matthew B. Collier, MS
Smain Lounici, MS
Anderson Orthopaedic Research Institute, c/o Rebecca Wolf, P.O. Box 7088, Alexandria, VA 22307

The authors did not receive grants or outside funding in support of their research or preparation of this manuscript. They did not receive payments or other benefits or a commitment or agreement to provide such benefits from a commercial entity. No commercial entity paid or directed, or agreed to pay or direct, any benefits to any research fund, foundation, educational institution, or other charitable or nonprofit organization with which the authors are affiliated or associated.

J Bone Joint Surg Am, 2002 Oct 01;84(10):1849-1855
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: Wear occurring at the interface between the polyethylene insert and metal baseplate of a modular tibial component has become an increasingly common finding at the time of revision total knee arthroplasty. Although this so-called backside wear on retrieved polyethylene inserts has been evaluated in prior studies, wear on retrieved metal baseplates has not been described, to our knowledge. The purposes of the present study were to characterize backside wear on retrieved polyethylene inserts and on the mating surfaces of their corresponding baseplates and to investigate if there is a relationship between backside wear and relative motion of the modular elements.

Methods: Twenty-nine retrieved modular tibial components of twelve fixed-bearing designs were analyzed in vitro with regard to backside wear and relative motion between the polyethylene insert and the metal baseplate. We graded the backside of each polyethylene insert and the mating surface of the metal baseplate for wear with use of a scoring system that consisted of three modes of wear and three levels of severity of wear. Relative motion between the insert and the baseplate was measured in the transverse plane with use of a mechanical testing machine. These measurements were used to compute the insert motion index, which served to quantify unrestricted motion of the insert with respect to the baseplate.

Results: The mean insert motion index for the tibial components was 416 µm (range, 104 µm to 760 µm). On a wear-grading scale ranging from 0 to 54 (with 0 indicating no wear), the mean backside wear score was 30 (range, 12 to 48) for the inserts and 28 (range, 7 to 51) for the baseplates. Insert motion was positively correlated with backside polyethylene wear (p = 0.003) and baseplate wear (p < 0.001). Baseplate wear was strongly correlated with backside polyethylene wear (p < 0.001).

Conclusions: Backside wear was correlated with the relative motion between the polyethylene insert and the metal baseplate. New locking mechanism designs directed toward better methods of securing the polyethylene insert to the tibial tray are needed to minimize the generation of particulate wear debris at the modular interface.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $35
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org

    References

    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe





    Related Content
    The Journal of Bone & Joint Surgery
    JBJS Case Connector
    Topic Collections
    Related Audio and Videos
    PubMed Articles
    Clinical Trials
    Readers of This Also Read...
    JBJS Jobs
    01/22/2014
    Pennsylvania - Penn State Milton S. Hershey Medical Center
    01/08/2014
    Pennsylvania - Penn State Milton S. Hershey Medical Center
    04/16/2014
    Georgia - Choice Care Occupational Medicine & Orthopaedics
    03/19/2014
    Virginia - VIRGINIA COMMONWEALTH UNIVERSITY MEDICAL CENTER